Holter ECG and BP monitor
CARDIOSENS — ECG and BP Holter
A digital 3-channel Holter ECG and a combined ECG+BP system. A 4G telemetry option is available.
Digital 3-channel Holter ECG
- 24–72 hours of ECG
- Miniature size, 60 g
- Only 4 wires for 3 leads
- Built-in battery and built-in 32 GB memory
Digital 3-channel Holter ECG + BP
- 24–72 hours of ECG
- 24 hours of BP readings
- Miniature size, up to 200 g
- Only 4 wires for 3 leads
- Built-in battery and built-in 32 GB memory
Key specifications
- ECG recording
- 24–72 hours, 3 channels
- Input signal range
- ± 8 mV
- Bandwidth
- 0.5–75 Hz
- CMRR
- 110 dB
- Internal noise
- < 5 µV
- Weight
- ≈ 60 g (ECG) · ≈ 200 g (ECG+BP)
- BP (BP version)
- up to ~250 readings; adult and pediatric modes
- BP range
- 10–280 mmHg (adults) · 5–155 mmHg (children)
- Memory
- built-in, 32 GB
- Interfaces
- USB-C, Bluetooth
Public procurement codes (Prozorro)
Ready-to-use wording for Ukrainian tender documents, specifications and contracts. We can send technical requirements on request.
Комплекс моніторингу електрокардіосигналів (амбулаторний реєстратор за методом Холтера)
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
35162Реєстратор амбулаторний для тривалого електрокардіографічного моніторингу- НК 031:2024
Z12050402Реєстратори ЕКГ за холтером
Комплекс комбінованого амбулаторного моніторингу електрокардіосигналів та артеріального тиску з функцією автоматичного вимірювання артеріального тиску
- ДК 021:2015
33120000-7Системи реєстрації медичної інформації та дослідне обладнання- НК 024:2023
35162Реєстратор амбулаторний для тривалого електрокардіографічного моніторингу- НК 031:2024
Z12050402Реєстратори ЕКГ за холтером

Software CARDIOSENS
- Patient file: multiple databases and folders, monitoring diary, network use
- Recorder programming and control ECG (USB-C / Bluetooth), record download to PC over USB
- Artifact filtering, QRS detection and markup
- Morphological classification (normal, SVE, VE, aberrant, unclassified, artifacts; with a pacemaker — spontaneous / paced / fusion), classifier training and manual cluster editing
- P waves, P–Q intervals, atrial activity; scatterplot; electronic magnifier
- Automatic rhythm and conduction analysis: counts per 24 h, hour, day and night; episode descriptions; dictionary and manual edits
- Rhythm: tachycardia, bradycardia, pauses; atrial and ventricular fibrillation / flutter
- SVE and VE: single, interpolated, couplets, polymorphic, bi-/tri-/quadrigeminy, runs; SVT and VT; R-on-T
- Conduction disorders and AV block; high-grade ectopy (Lown 4b, 5); automatic pacemaker-pulse detection
- ST-T with decision-rule and episode-boundary correction; QRS morphology; Q–T intervals
- HRV in time and frequency domains (ESC working-group norms, PARS); AC/DC via PRSA; heart-rate turbulence (HRT)
- EASI-12 reconstruction of 12 leads from the 3-channel recording
- Auto report with ECG strips, conclusion-form editor, print
- Export of records and tables (network, e-mail, Excel, MATLAB)
- HIS integration
- Integration with our own shared-access and data-storage system CLORM
- Interface, reports and printouts in Ukrainian, English and Russian
ECG+BP version
- BP per NBREP: SBP / DBP / mean / pulse pressure and HR trends; cuff oscillation and pulse-wave review
- Distribution histograms, diagrams, correlation, cosinor analysis
- Pressure-load indices (measurement, time, area); BP variability (VAR1–3)
- Nocturnal dip (SNS), day / night / custom intervals; morning-surge magnitude and rate
- Calibration BP readings and offsets; adult and pediatric protocols
- Summary tables and a separate BP report (form editor, fragment and full-report print)
Request a quote CARDIOSENS · detailed brochure · PDF 9.26 MB